Quick and easy calibration of vibration sensors and vibration measurement systems for acceleration, velocity and displacement
Removable clamping device for dynamic calibration of non-contact (inductive / Eddy Current) displacement transducers at 159.2 Hz with threaded adapters for all common sensor types
Clock output for synchronisation of frequency-selective measuring systems, e.g. as “keyphasor” signal in machine monitoring systems
Load-independent vibration level
Switchable oscillation frequencies and amplitudes
Battery-powered for mobile use
Robust construction
Note: At the vibration frequency of 159.2 Hz, an acceleration of 1, 2, 5, 10 or 20 m/s² results in a vibration velocity of 1, 2, 5, 10 or 20 mm/s and a vibration displacement of 1, 2, 5, 10 or 20 µm (effective values in each case).
Technische Daten
Technical data
Vibration amplitude | 1 / 2 / 5 / 10 / 20 m/s² |
Vibration frequency | 15,92 / 40 / 80 / 159,2 / 320 / 640 / 1280 Hz |
Sensor mass (max.) | 50 .. 500 g (depending on frequency and amplitude) |